Complementary
Absolute accuracy error
+/- 6 °C at 25 °C for thermocouple R, S 0...200 °C
+/- 0.4 % of full scale at 25 °C for thermocouple K - 200...0 °C
+/- 0.4 % of full scale at 25 °C for thermocouple J - 200...0 °C
+/- 0.4 % of full scale at 25 °C for thermocouple E - 200...0 °C
+/- 0.4 % of full scale at 25 °C for thermocouple T - 200...0 °C
+/- 0.4 % of full scale at 25 °C for thermocouple N - 200...0 °C
+/- 0.1 % of full scale at 25 °C for analogue input voltage/current
+/- 0.1 % of full scale at 25 °C for thermocouple C 0...2315 °C
+/- 0.1 % of full scale at 25 °C for Pt 100/Pt 1000, Ni 100/ Ni 1000 temperature probe
+/- 0.1 % of full scale at 25 °C for thermocouple R, S 200...1760 °C
+/- 0.1 % of full scale at 25 °C for thermocouple B 300...1820 °C
+/- 0.1 % of full scale at 25 °C for thermocouple K 0...1300 °C
+/- 0.1 % of full scale at 25 °C for thermocouple J 0...1000 °C
+/- 0.1 % of full scale at 25 °C for thermocouple E 0...800 °C
+/- 0.1 % of full scale at 25 °C for thermocouple T 0...400 °C
+/- 0.1 % of full scale at 25 °C for thermocouple N 0...1300 °C
+/- 0.1 % of full scale at 25 °C for analogue output voltage/current
range compatibility
Modicon M221
Modicon M241
Modicon M251
Modicon M262
Analogue output resolution
12 bits
Handle front plate colour
90 mm
Product or component type
Input/output analog module
[us] rated supply voltage
24 V DC
Current consumption
55 mA at 24 V DC via external supply no load
55 mA at 5 V DC via bus connector no load
60 mA at 5 V DC via bus connector full load
80 mA at 24 V DC via external supply full load
marking
CE
Analogue input number
2
Analogue input type
current 4...20 mA
current 0...20 mA
voltage 0...10 V
voltage - 10...10 V
thermocouple - 200...1000 °C with thermocouple J
thermocouple - 200...1300 °C with thermocouple K
thermocouple 0...1760 °C with thermocouple R
thermocouple 0...1760 °C with thermocouple S
thermocouple 0...1820 °C with thermocouple B
thermocouple - 200...400 °C with thermocouple T
thermocouple - 200...1300 °C with thermocouple N
thermocouple - 200...800 °C with thermocouple E
thermocouple 0...2315 °C with thermocouple C
Ni 100/Ni 1000 temperature probe - 60...180 °C
Pt 100 temperature probe - 200...850 °C
Pt 1000 temperature probe - 200...600 °C
Analogue output number
1
width (w)
23.6 mm
Analogue output type
Current: 4...20 mA
Current: 0...20 mA
Voltage: 0...10 V
Voltage: - 10...10 V
Load impedance ohmic
1 kOhm voltage
300 Ohm current
supply voltage limits
20.4…28.8 V
Local signalling
1 LED (green) for PWR
temperature drift
+/- 0.006 %FS/°C
Net weight
0.1 kg
Immunity to microbreaks
10 ms
input impedance
<= 50 Ohm current
>= 1 MOhm voltage
>= 1 MOhm thermocouple
>= 1 MOhm temperature probe
Mounting support
Top hat type TH35-15 rail conforming to IEC 60715
Top hat type TH35-7.5 rail conforming to IEC 60715
plate or panel with fixing kit
depth
70 mm
Load type
Resistive
repeat accuracy
+/-0.5 %FS for input
+/- 0.4 %FS for output
Pollution degree
2
electrical connection
11 x 2.5 mm² removable spring terminal block with pitch 5.08 mm adjustment for inputs, outputs and supply
Non-linearity
+/- 0.01 %FS, analogue output type: analog output
+/- 0.1 %FS, analogue input type: analog input
Output ripple
20 mV
Permissible continuous overload
13 V, analogue input type: voltage
40 mA, analogue input type: current
Insulation
Between input and supply at 1500 V AC
Between input and internal logic at 500 V AC
Between output and supply at 1500 V AC
Between output and internal logic at 500 V AC
Stabilisation time
1 ms
Lsb value
0.15 mV 0...10 Vvoltage
0.30 mV - 10...10 Vvoltage
0.30 µA 0...20 mAcurrent
0.244 µA 4...20 mAcurrent
0.1 °Ctemperature probe
0.1 °Cthermocouple
2.44 mV 0...10 V voltage
4.88 mV - 10...10 V voltage
4.88 µA 0...20 mA current
3.91 µA 4...20 mA current
Resistance to electrostatic discharge
8 kV in air conforming to EN/IEC 61000-4-2
4 kV on contact conforming to EN/IEC 61000-4-2
Shock resistance
15 gn for 11 ms
conversion time
100 ms + 100 ms per channel + 1 controller cycle time for analogue input thermocouple
100 ms + 100 ms per channel + 1 controller cycle time for analogue input temperature probe
10 ms + 10 ms per channel + 1 controller cycle time for analogue input voltage/current
Resistance to electromagnetic fields
10 V/m 80 MHz...1 GHz conforming to EN/IEC 61000-4-3
3 V/m 1.4 GHz...2 GHz conforming to EN/IEC 61000-4-3
1 V/m 2 GHz...3 GHz conforming to EN/IEC 61000-4-3
Surge withstand
1 kV power supply common mode conforming to EN/IEC 61000-4-5
0.5 kV power supply differential mode conforming to EN/IEC 61000-4-5
1 kV I/O common mode conforming to EN/IEC 61000-4-5
0.5 kV I/O differential mode conforming to EN/IEC 61000-4-5
Sampling duration
10 ms, analogue input type: voltage/current
100 ms, analogue input type: voltage/current
100 ms, analogue input type: thermocouple
100 ms, analogue input type: temperature probe
Type of cable
Twisted shielded pairs cable <30 m for input/output circuit
Environment
Ambient air temp for storage/transportation
-25…70 °C
ip degree of protection
IP20
operating altitude
0...2000 m
Range of product
Modicon TM3
ambient air temperature for operation
-10…55 °C horizontal installation
-10…35 °C vertical installation
Resistance to conducted disturbances
10 V 0.15...80 MHz conforming to EN/IEC 61000-4-6
3 V spot frequency (2, 3, 4, 6.2, 8.2, 12.6, 16.5, 18.8, 22, 25 MHz) conforming to Marine specification (LR, ABS, DNV, GL)
standards
EN/IEC 61131-2
EN/IEC 61010-2-201
Storage altitude
0…3000 m
relative humidity
10…95 %, without condensation (in operation)
10…95 %, without condensation (in storage)
electromagnetic emission
Radiated emissions - test level: 40 dBμV/m QP class A ( 10 m) at 30…230 MHz conforming to EN/IEC 55011
Radiated emissions - test level: 47 dBμV/m QP class A ( 10 m) at 230…1000 MHz conforming to EN/IEC 55011
vibration resistance
3.5 mm at 5…8.4 Hz on DIN rail
3 gn at 8.4…150 Hz on DIN rail
resistance to fast transients
1 kV (I/O) conforming to EN/IEC 61000-4-4
resistance to magnetic fields
30 A/m conforming to EN/IEC 61000-4-8